4. Calibration
page 34
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
Waiting time 10 sec.
End of Calibration
If the last calibration standard has been injected, press F1- End. Now the calibration function is
calculated automatically. An outlier is marked as a square. It can be accepted to be included in
the calibration function (F1- Yes) or it can be excluded from the calibration function
(F2- No).
If the outlier is not accepted (F2- No), the user is asked to repeat the calibration point.
On a second page calibration data are
displayed.