C-3 Measuring DC Current
1. Connect the black test lead to COM jack .
When measuring below 200mA, connect the
red test lead to mA jack. When measuring
20A or below, connect the red test lead to
20A jack.
2. Set the rotary switch to A
3. Connect the test leads in series with the object
to be measured, the LCD display the measuring
value and polarity of red test lead.
“ ”
“ ”
“ ”
“ ” .
Caution
C-4 Measuring AC Current
1. Connect the black test lead to COM jack.
When measuring below 200mA, connect the
red test lead to “mA” jack. When measuring
20A or below, connect the red test lead to
20A jack.
2. Set the rotary switch to A~ .
3. Connect the test leads in series with the object
to be measured.
“ ”
“ ”
“ ”
1) Please refer to DC Current Caution 1, 2, 3.
Caution
1) If magnitude of the current is unknown, always
start with the highest range and reduce until
satisfactory reading is obtained.
2) If 1 is shown on the LCD, which means the
Meter is overloaded, then set the measuring
range to higher.
3)
means the maximum input current is 200
mA, overload will cause the burn of fuse. 20A
range does not have fuse protection
“ ”
“ ”
“ ”
11
Model UT56: OPERATING MANUAL