UNI-T UT3000 series Instrucciones de operaciones Pagina 20

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 98
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 19
3. Operating instructions
page 20
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
F
IGURE
:
S
IGNAL
P
EAK DURING THERMAL DESORPTION OF THE
G
OLD
T
RAP
.
Sampling
Zeroing
Heating
Trap Cooling
Purge Cell
Trap
Cleaning
Trap Cooling
UT-3000
Sampling
Zeroing
Heating Trap Cooling
Purge Cell
Trap
Cleaning Cooling
UT-3000 ng (Tedlar Version only)
Cleaning
Cooling
F
IGURE
:
UT-3000:
MEASUREMENT CYCLE
Sampling Zeroing
Heating / Cleaning Cooling
5 s to 15 min; depending on
sample volume
11 s 20 s 45 s for sample vol. 1 L and
up
>45 s for sample vol < 1 l
T
ABLE
:
DURATIONS OF THE ANALYSIS STEPS
Repeat Zero: Zeroing is repeated up to 10 times if the heating flow is out of range
(7 l/h < heating flow <13 l/h)
Vista de pagina 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 97 98

Comentarios a estos manuales

Sin comentarios