UNI-T UT3000 series Instrucciones de operaciones Pagina 46

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 98
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 45
4. Calibration
page 46
Mercury Instruments
UT-3000-E_V2.11.docx
31.07.12
Analytical Technologies
4.5.5 Manual Start of the Automatic Calibration
F3- Start automatic Calibration is selected in the Calibration Mode menu. Now an automatic
calibration is immediately started using the concentrations entered under F1-Cal. Concentration.
The following screens can be noticed consecutively:
After that the standard cleaning routine of the GoldTrap is done:
Trap Cleaning / Trap Cooling / Zeroing / Trap Cleaning / Trap Cooling / Zeroing /
Trap Cleaning / Trap Cooling
Now the routine for each calibration point is starting and repeated as often as your amount of
different concentrations.
Vista de pagina 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 97 98

Comentarios a estos manuales

Sin comentarios